Defining E1 and E2 Weight Classes
E1 Test Weights and E2 are accuracy classes for reference weights defined by the International Organization of Legal Metrology (OIML) in recommendation R111. E1 represents the highest accuracy class typically used for calibrating other high-precision weights, while E2 is one step below, often used for calibrating F1 class weights or high-precision analytical balances.
The primary distinction lies in their maximum permissible errors (MPE). For a 1 kg weight, the MPE for an E1 class weight is ±0.5 mg, whereas for an E2 class weight, it is ±1.6 mg. This tenfold difference in precision dictates their specific roles in the traceability chain of mass measurement.
Tolerance Limits and Technical Specifications
Understanding the specific tolerance limits is crucial for selecting the right weight for your calibration needs. The OIML R111 standard specifies these limits based on the nominal value of the weight.
| Nominal Value | E1 Class MPE (mg) | E2 Class MPE (mg) |
|---|---|---|
| 1 mg | 0.003 | 0.010 |
| 1 g | 0.025 | 0.080 |
| 100 g | 0.16 | 0.50 |
| 1 kg | 0.50 | 1.60 |
| 10 kg | 2.50 | 8.00 |
These tight tolerances require that E1 and E2 weights be manufactured from high-density materials, such as stainless steel or nickel-chromium alloys, to minimize surface area and reduce the effects of air buoyancy and adsorption.

Primary Applications in Metrology
The choice between E1 and E2 depends largely on the position of the weight within the calibration hierarchy and the required uncertainty of the measurement.
E1 Class Applications
E1 weights are primarily used by national metrology institutes and accredited calibration laboratories. Their main role is to calibrate E2 class reference weights. They are also used for the direct calibration of high-precision comparators and balances with readability of 0.1 mg or better.
E2 Class Applications
E2 weights serve as reference standards in industrial calibration labs. They are commonly used to calibrate F1 class working weights and high-accuracy analytical balances. E2 weights are the standard choice for pharmaceutical and chemical industries where precise mass measurement is critical for formulation and quality control.
Handling and Maintenance Best Practices
Maintaining the accuracy of E1 and E2 weights requires strict adherence to handling protocols. Even minor contamination can cause a weight to exceed its permissible error limits.
- Use Gloves: Always handle weights with clean, lint-free gloves or specialized tweezers to prevent oil and sweat transfer from skin.
- Control Environment: Store and use weights in a temperature-controlled environment (typically 20°C ± 2°C) to minimize thermal expansion effects.
- Regular Calibration: E1 weights should be recalibrated annually, while E2 weights may have a longer interval depending on usage frequency, but typically no more than two years.
- Cleaning Procedures: Use only approved cleaning agents, such as pure ethanol or acetone, and allow weights to acclimatize to room temperature before use after cleaning.
Key Differences Between E1 and E2
While both classes represent high-precision standards, several factors differentiate them beyond just tolerance limits.
- Cost: E1 weights are significantly more expensive due to the rigorous manufacturing and adjustment processes required to achieve tighter tolerances.
- Material Density: E1 weights often require higher density materials (≥ 8000 kg/m³) to reduce air buoyancy uncertainty, whereas E2 allows for slightly lower densities.
- Magnetic Susceptibility: E1 weights must have extremely low magnetic susceptibility to avoid interference with electromagnetic force compensation balances, a requirement that is less stringent for E2.
- Traceability: E1 weights are usually calibrated directly against national primary standards, while E2 weights are often calibrated against E1 references.